m1000 Design

Overall Dimensions
 Width (cm/in.)Depth (cm/in.) Height (cm/in.) 
Cabinet142.9 cm / 56.3 in79.4 cm / 31.3 in76.2 cm 30.0 in
m1000 instrument 142.9 cm / 56.3 in 76.8 cm / 30.3 in (127.0 cm / 50.0 in)d 
 (176.5 cm / 69.5 in)b (124.5 cm / 49.0 in)c  
a) Dimensions do not include computer, monitor or printer
b) Side service door open
c) Front safety panel open
d) RoMa extended


Instrument Weight 
Weight 
Cabinet103.4 kg / 228.0 lbs
m1000 instrument 178.3 kg / 393.0 lbse 

e) Weight includes accessories packed inside the cabinet and instrument

Electrical
  
Circuit breaker15 A
Equipment and safety classI
Frequency50-60 Hz 
Groundingproperly grounded electrical outlet 
Power1200 VA 
Line voltage 100-240 V AC 
Number of electrical outlets minimum 4(instrument, PC, monitor, printer) 





Performance
 Liquid Handling Precision
TipsDiTi 200 and 100079.4 cm / 31.3 in
Pipetting volume range10-1000 ul 76.8 cm / 30.3 in 
Precisionf at 10 ul dispense volume
at 100 ul dispense volume
CV < 3 %
CV < 0.5 %
f) Values are only achieved if maintenance instructions and schedule have strictly been followed

Available in US